Key facts
| Recalling firm | Wipro GE Healthcare Private Ltd. |
|---|---|
| Manufacturer | Not disclosed in the FDA record. |
| Brand name | Not disclosed in the FDA record. |
| Product description | GE Giraffe Warmer, Model/Catalog Number M1118179-01090979. Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. |
| Classification | Class II |
| Recall number | Z-3080-2024 |
| Report date | Not disclosed in the FDA record. |
| Status | Ongoing |
| Reason for recall | GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits. |
What consumers should do
Facilities, clinicians, and consumers holding an affected unit or lot of GE Giraffe Warmer, Model/Catalog Number M1118179-01090979. Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. should stop using it for the purpose described in the recall notice and hold remaining stock pending instructions from Wipro GE Healthcare Private Ltd.. FDA recall records typically direct the recalling firm to notify known accounts directly with a recall letter describing the specific corrective action -- replacement, field correction, refund, or return. Anyone who received the product through a distributor rather than directly from Wipro GE Healthcare Private Ltd. should confirm with their supplier whether their specific lot or unit is on the affected list, since recall notices are lot- or unit-specific and not every unit of a product line is necessarily included. Questions about a specific unit's status should go to Wipro GE Healthcare Private Ltd. directly or to the distributor that supplied it; Argus HQ is a public-record aggregator, not the recalling firm, and cannot confirm individual unit status.
What this classification means
A Class II recall is FDA’s middle classification. FDA assigns Class II when use of, or exposure to, the recalled product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote. FDA still requires the recalling firm to notify affected accounts and submit status reports on the correction, though the monitoring intensity is lower than for a Class I recall. FDA's Center for Devices and Radiological Health (for devices) or Center for Drug Evaluation and Research (for drugs) reviews the classification and can require additional corrective action if the initial response is judged insufficient. Because this record is filed as Class II, professionals and consumers handling an affected lot should treat FDA's own classification -- not this page's summary of it -- as the authoritative risk signal. A Class I recall is FDA’s most serious classification. FDA assigns Class I when there is a reasonable probability that use of, or exposure to, the recalled product will cause serious adverse health consequences or death. This classification does not mean harm has already occurred in every case -- it reflects FDA’s assessment of the realistic worst-case risk profile of the defect described in the recall, based on the reason for recall reported by the firm. Class I recalls receive the closest FDA monitoring of the three classes: the recalling firm typically must submit a recall strategy, periodic status reports, and effectiveness checks confirming that the affected lots have actually been removed from distribution or corrected in the field.
Wipro GE Healthcare Private Ltd.’s FDA history
Argus HQ has recorded 10 total FDA actions tied to Wipro GE Healthcare Private Ltd.: 0 warning letters, 10 recalls, 0 approval records, and 0 Form 483 inspection citations.
Frequently asked questions
- Who should I contact about this recall?
- Contact Wipro GE Healthcare Private Ltd. directly, or the distributor/pharmacy that supplied the product, to confirm whether your specific lot or unit is affected. Argus HQ aggregates the public FDA recall record and is not the recalling firm.
- Has Wipro GE Healthcare Private Ltd. had other FDA recalls or enforcement actions?
- Yes. Argus HQ has recorded 10 total FDA actions tied to Wipro GE Healthcare Private Ltd., including 10 recalls and 0 warning letters.
- What is the FDA recall number for this event?
- FDA recall number Z-3080-2024.
- What is the current status of this recall?
- Status as recorded by FDA: Ongoing.
Know if this happens again — we'll email you.
Free. No spam, just recall alerts for what you tell us to watch.
Related enforcement actions
Full FDA history for Wipro GE Healthcare Private Ltd.- Recall: Wipro GE Healthcare Private Ltd. — Class II recall: GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software v… →
- Recall: Wipro GE Healthcare Private Ltd. — Class II recall: GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software v… →
- Recall: Wipro GE Healthcare Private Ltd. — Class II recall: GE Panda iRES Warmer, Model/Catalog Numbers: 1) M1112198-01045518; 2) M1112… →
- Recall: Wipro GE Healthcare Private Ltd. — Class I recall: GE Healthcare Care Plus Hood Door FRU kit, Part Numbers: 6600-0082-850, 6600-050… →
- Recall: Wipro GE Healthcare Private Ltd. — Class I recall: GE Healthcare Care Plus Hood Field Replacement Unit (FRU) kit, Part Numbers: 66… →
- Recall: Wipro GE Healthcare Private Ltd. — Class I recall: GE Healthcare Lullaby Incubator Model Hood Door FRU kit, Part Numbers: M1169767… →
- Recall: Wipro GE Healthcare Private Ltd. — Class I recall: GE Healthcare neonatal incubator, models: a) Lullaby Incubator, b) Lullaby… →
- Recall: Wipro GE Healthcare Private Ltd. — Class I recall: GE Healthcare neonatal incubator, models: a) Care Plus, b) Care Plus 1000,… →
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
Argus HQ Research (2024). FDA Class II Recall: GE Giraffe Warmer, Model/Catalog Number M1118179-01090979. Infant radiant… by Wipro GE Healthcare Private Ltd.. Digital Empire LLC. Retrieved from https://argushq.ai/recall/wipro-ge-healthcare-private-z-3080-2024-2024-10-02
"FDA Class II Recall: GE Giraffe Warmer, Model/Catalog Number M1118179-01090979. Infant radiant… by Wipro GE Healthcare Private Ltd.." Argus HQ Research, Digital Empire LLC, 2024, argushq.ai/recall/wipro-ge-healthcare-private-z-3080-2024-2024-10-02.
Argus HQ Research. "FDA Class II Recall: GE Giraffe Warmer, Model/Catalog Number M1118179-01090979. Infant radiant… by Wipro GE Healthcare Private Ltd.." Digital Empire LLC. Accessed July 17, 2026. https://argushq.ai/recall/wipro-ge-healthcare-private-z-3080-2024-2024-10-02.
@misc{argushq_argushq_ai_recall_wipro_ge_healthcare_private_z_3080_2024_2024_10_02_2024,
title = {FDA Class II Recall: GE Giraffe Warmer, Model/Catalog Number M1118179-01090979. Infant radiant… by Wipro GE Healthcare Private Ltd.},
author = {{Argus HQ Research}},
year = {2024},
publisher = {Digital Empire LLC},
url = {https://argushq.ai/recall/wipro-ge-healthcare-private-z-3080-2024-2024-10-02},
note = {Accessed: July 17, 2026}
}Source: FDA.gov — Recalls ↗